MODEL
|
HV-85
|
HV-110
|
TEMPERATURE |
|
Operating Range |
105ºC to 135ºC
|
Display Range |
137ºC
|
Warming Range |
45ºC to 60ºC
|
Agar Dissolution |
Heating at 60ºC to 100ºC for up to 1 hour (temperature and time programmable), followed by 45ºC to 60ºC (programmable for up to 20 hours to prevent coagulations)
|
GAUGE PRESSURE |
|
Operating Range |
3 to 30 psi (0.021 to 0.206MPa)
|
Maximum Allowable |
38 psi ( 0.26MPa)
|
TIMER |
|
Sterilization |
1 to 250 minutes
|
Delayed Auto-start |
Program autoclave to start 1 minute to 7 days later
|
DIMENSIONS |
|
Overall (W x D x H) |
26.0" x 25.6" x 39.4" (66 cm x 65 cm x 100 cm)
|
26.0" x 25.6" x 46.5" (66 cm x 65 cm x 118 cm)
|
Chamber (Diameter x Height) |
85 liters 16.5" x 24.2" (42.0 cm x 61.5 cm)
|
110 liters 16.5" x 31.3" (42.0 cm x 79.5 cm)
|
SAFETY DEVICES |
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker
|
PROGRAMS |
|
1. For agar |
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Warming -> Completion
|
2. For liquids |
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Completion
|
3. For solids |
Heating -> Sterilization -> Exhaust (fast, default) -> Completion
|
4. For melting agar or for faster start-up |
Rapid melting at 60°C to 100°C (temperature and time programmable) -> Maintain at 45°C to 60°C (temperature programmable) -> Completion
|
ACCESSORIES |
|
Stainless steel |
Mesh wire baskets, buckets (pails), and round test tube rack
|
Factory or field installed |
Floating (load) sensor, digital printer, and data logger
|
POWER |
|
Source |
208V/220V/240V, 50/60 Hz, single phase
|
Consumption |
3000 watts (3000VA)
|
4000 watts (4000VA)
|
SHIPPING WEIGHT |
282 lb (128 kg)
|
308 lb (140 kg)
|